RobotRisk › Asensus Surgical › 15 Sep 2023
Class I recall of Senhance surgical robot over unintended arm movement
S3 · 15 Sep 2023 · Asensus Surgical
Record
| Company | Asensus Surgical |
| Industry | Medical robotics |
| Type(s) | recall |
| Date | 15 Sep 2023 (day precision) |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | (recall, enforcement, or fleet-wide fault) |
| Scale | 5 devices in the US, distributed April-September 2023 |
| Confidence | high |
| Verified flag | Cited source on file; not independently re-verified |
Summary
FDA classified Asensus Surgical's recall of the Senhance Surgical System as Class I after malfunctions caused uncontrolled, continuous rotation of the laparoscope instrument actuator arm even after the surgeon disengaged teleoperation, creating potential for critical tissue trauma; Asensus reported no patient harm to date and instructed customers to stop use until a software update (v2.7.5) was applied.
Primary source
Publisher: FDA · Retrieved: 2026-09-16
Supporting source rows (1)
| Date | Supporting source row | Publisher | Link |
|---|---|---|---|
| 15 Sep 2023 | Class I recall of Senhance surgical robot over unintended arm movement | FDA | FDA |
Medical robotics context
recall context
Asensus Surgical industry position
Related incidents
Other Asensus Surgical entries
- None on file.
Report a correction
Also from Critical Systems Analysis: CSA - functional safety engineering · Company directory · FSTalent - functional safety jobs