RobotRisk Autonomous Systems Incident & Risk Register
RobotRiskAsensus Surgical › 15 Sep 2023

Class I recall of Senhance surgical robot over unintended arm movement

S3 · 15 Sep 2023 · Asensus Surgical

Record

CompanyAsensus Surgical
IndustryMedical robotics
Type(s)recall
Date15 Sep 2023 (day precision)
Time of daynot documented
Locationno physical site (recall / fleet-wide / aggregate record)
Severity (recall, enforcement, or fleet-wide fault)
Scale5 devices in the US, distributed April-September 2023
Confidencehigh
Verified flagCited source on file; not independently re-verified

Summary

FDA classified Asensus Surgical's recall of the Senhance Surgical System as Class I after malfunctions caused uncontrolled, continuous rotation of the laparoscope instrument actuator arm even after the surgeon disengaged teleoperation, creating potential for critical tissue trauma; Asensus reported no patient harm to date and instructed customers to stop use until a software update (v2.7.5) was applied.

Primary source

FDA

Publisher: FDA · Retrieved: 2026-09-16

Supporting source rows (1)
DateSupporting source rowPublisherLink
15 Sep 2023 Class I recall of Senhance surgical robot over unintended arm movement FDA FDA