RobotRisk Autonomous Systems Incident & Risk Register
RobotRiskCMR Surgical › 09 Jun 2026

Class II recall of Versius surgeon console over disabled Secure Boot cybersecurity control

S2 · 09 Jun 2026 · CMR Surgical

Record

CompanyCMR Surgical
IndustryMedical robotics
Type(s)recall
Date09 Jun 2026 (day precision)
Time of daynot documented
Locationno physical site (recall / fleet-wide / aggregate record)
Severity (minor injury or single-unit damage)
Scale2 units distributed in the US (California)
Confidencehigh
Verified flagCited source on file; not independently re-verified

Summary

CMR Surgical recalled Versius Surgical System surgeon consoles (model V-SS-1020) after determining Secure Boot was mistakenly not enabled at manufacturing time, creating a potential cybersecurity risk; the firm notified customers, committed to reconfiguring affected consoles, and instructed users to inspect systems and monitor tamper-evident labels for unauthorized access.

Primary source

FDA

Publisher: FDA · Retrieved: 2026-09-16

Supporting source rows (1)
DateSupporting source rowPublisherLink
09 Jun 2026 Class II recall of Versius surgeon console over disabled Secure Boot cybersecurity control FDA FDA

Medical robotics context

recall context

CMR Surgical industry position

Related incidents

Other CMR Surgical entries

Report a correction

Open a prefilled GitHub issue

Also from Critical Systems Analysis: CSA - functional safety engineering · Company directory · FSTalent - functional safety jobs