FDA Class II recall: Galaxy System (GAL-001).

S3 · 28 Oct 2024 · Noah Medical

Record

CompanyNoah Medical
IndustryMedical robotics
Type(s)recall
Date28 Oct 2024 (day precision)
Time of daynot documented
Locationno physical site (recall / fleet-wide / aggregate record)
Severity (recall, enforcement, or fleet-wide fault)
ScaleGalaxy System (GAL-001).
Confidencehigh
Verified flagYes - independently re-checked against the cited source

Summary

FDA enforcement record (Class II, recall Z-0742-2025): Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increa

Primary source

openFDA enforcement

Publisher: openFDA enforcement · Retrieved: 2026-09-16

Supporting source rows (1)
DateSupporting source rowPublisherLink
28 Oct 2024 FDA Class II recall: Galaxy System (GAL-001). openFDA enforcement openFDA enforcement