FDA Class II recall: Galaxy System (GAL-001).
S3 · 28 Oct 2024 · Noah Medical
Record
| Company | Noah Medical |
| Industry | Medical robotics |
| Type(s) | recall |
| Date | 28 Oct 2024 (day precision) |
| Time of day | not documented |
| Location | no physical site (recall / fleet-wide / aggregate record) |
| Severity | (recall, enforcement, or fleet-wide fault) |
| Scale | Galaxy System (GAL-001). |
| Confidence | high |
| Verified flag | Yes - independently re-checked against the cited source |
Summary
FDA enforcement record (Class II, recall Z-0742-2025): Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increa
Primary source
Publisher: openFDA enforcement · Retrieved: 2026-09-16
Supporting source rows (1)
| Date | Supporting source row | Publisher | Link |
|---|---|---|---|
| 28 Oct 2024 | FDA Class II recall: Galaxy System (GAL-001). | openFDA enforcement | openFDA enforcement |