RobotRiskGlobus Medical › 14 Feb 2024

FDA Class II recall: ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes

S3 · 14 Feb 2024 · Globus Medical

Record

CompanyGlobus Medical
IndustryMedical robotics
Type(s)recall
Date14 Feb 2024 (day precision)
Time of daynot documented
Locationno physical site (recall / fleet-wide / aggregate record)
Severity (recall, enforcement, or fleet-wide fault)
ScaleExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball
Confidencehigh
Verified flagYes - independently re-checked against the cited source

Summary

FDA enforcement record (Class II, recall Z-1007-2024): Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-con

Primary source

openFDA enforcement

Publisher: openFDA enforcement · Retrieved: 2026-09-16

Supporting source rows (1)
DateSupporting source rowPublisherLink
14 Feb 2024 FDA Class II recall: ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes openFDA enforcement openFDA enforcement

Globus Medical other incidents timeline

Medical robotics context

recall context

Globus Medical industry position